CALIBRATION PROCEDURE
NI PXIe-4154
This document contains information for calibrating the NI PXIe-4154 Battery Simulator. For
more information about calibration, visit
ni.com/calibration
.
Contents
Software Requirements............................................................................................................. 1
Documentation Requirements .................................................................................................. 2
Password................................................................................................................................... 2
Calibration Interval................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 4
Calibration Procedures ............................................................................................................. 5
Voltage Output Verification ..................................................................................... 6
Current Output Verification...................................................................................... 7
Voltage Measurement Verification .......................................................................... 9
Current Measurement Verification........................................................................... 10
Output Resistance Verification (Channel 0 Only).................................................... 14
Load Regulation Verification ................................................................................... 17
Remote Sense Accuracy Verification....................................................................... 19
Transient Response Verification (Channel 0 Only) ................................................. 20
Considerations .......................................................................................................... 24
Adjusting Current Measurement .............................................................................. 25
Adjusting Voltage Output......................................................................................... 26
Adjusting Current Output ......................................................................................... 27
Adjusting Voltage Measurement .............................................................................. 29
Adjusting Output Resistance (Channel 0 Only) ....................................................... 30
Where to Go for Support .......................................................................................................... 31
Software Requirements
Calibrating the NI PXIe-4154 requires installing NI-DCPower version 1.5 or later on a Windows
calibration system. You can download the NI-DCPower instrument driver from
ni.com/
drivers
. NI-DCPower supports programming an External Calibration in the C and LabVIEW
application development environments (ADEs). When you install NI-DCPower, you only need
to install support for the ADE that you intend to use.