NI PXIe-4154 Calibration Procedure
|
© National Instruments Corporation
|
13
0
3 A
100 m
Ω
-3 A
0.14% +
0.7 mA
-2.25 A
0.1025% +
0.7 mA
-1.5 A
0.065% +
0.7 mA
-750 mA
0.04% +
0.7 mA
0 A
0.04% +
0.7 mA
750 m
0.04% +
0.7 mA
1.5 A
0.065% +
0.7 mA
2.25 A
0.1025% +
0.7 mA
3 A
0.14% +
0.7 mA
1
1.5 A
100 m
Ω
0 A
0.12% +
0.35 mA
375 mA
0.12% +
0.35 mA
750 mA
0.17% +
0.35 mA
1.125 A
0.245% +
0.35 mA
1.5 A
0.32% +
0.35 mA
Table 5.
NI PXIe-4154
Current Measurement Verification Points (Continued)
Channel
Range
Precision
Shunt
Value
Test
Point
Measured
Current
Test
Limit
Reported
Current
As
Found
As
Left
As
Found
As
Left