NI PXIe-4154 Calibration Procedure
|
© National Instruments Corporation
|
31
After all test points are measured for a range, call the niDCPower Cal Adjust Output Resistance
VI or the
niDCPower_CalAdjustOutputResistance
function to calculate updated
calibration coefficients. Refer to Table 26 for parameters specific to adjustment applications for
this VI. Refer to the
NI DC Power Supplies and SMUs Help
for more information about
programming references common to VIs within the NI-DCPower API.
Figure 16.
Figure 15. NI-DCPower Cal Adjust Output Resistance VI Block Diagram
After the new coefficients are calculated, commit them to the hardware using the process
described in the
section.
Revision History
Table 25.
NI PXIe-4154
Output Resistance Adjustment Points
Channel
V Range
V Level
I Limit
Required
Adjustment Points
0
6 V
4 V
3 A
0
Ω
0.5
Ω
0.9
Ω
Table 26.
NI-DCPower Cal Adjust Output Resistance VI Parameter Descriptions
Parameter
Description
range
The range to be adjusted with these settings.
requested outputs
Array of requested output resistance test points required for adjustment.
For example, this would be 0
Ω
, 0.5
Ω
, and 0.9
Ω
for Channel 0.
measured outputs
Array of measurements made by external DMMs corresponding to
requested outputs
.
number of measurements
The number of elements in
requested outputs
and
measured output
s.
Revision
Section
Changes
373092C-01
August 2020
Transient Response
Verification (Channel 0 Only)
Removed test steps for Normal
Transient Response.
Adjusting Output Resistance
(Channel 0 Only)
Updated recommended resistance
values for NI-DCPower Cal Adjust
Output Resistance VI requested
output parameter.