NI PXIe-4154 Calibration Procedure
|
© National Instruments Corporation
|
21
Figure 9.
Channel 0 Transient Response Connection Diagram
3.
Configure the electronic load in current control dynamic mode with the settings listed in the
following table.
4.
Configure the oscilloscope to acquire a waveform according to the settings listed in the
following table.
5.
Configure the NI PXIe-4154 according to the settings listed in the following table.
6.
Measure the transient response recovery time and transient response dip, as displayed in
Figure 10. The transient response recovery time is measured from the time the dip begins
until the time at which the waveform crosses above the final voltage minus 20 mV.
Table 13.
Electronic Load Settings
Load High Current
Load Low Current
Load Period
Load Slew Rate
1.5 A
0.1 A
1 ms
200 mA/µs
Table 14.
Oscilloscope Settings
Volts/
Division
Time/
Division
Trigger
Slope
Trigger
Level
Input
Impedance
Input
Coupling
Sampling
Rate
Trigger
50 mV
20 µS
Negative
Edge
-20 mV
1 M
Ω
AC
≥
200 MS/s
Edge
Trigger
Table 15.
NI PXIe-4154 Settings
V Range
I Range
V Level
I Limit
6 V
3 A
6 V
3 A
NI PXIe-4154
Electronic
Load
+
–
+
–
Sense –
Sense +
Output –
Digital
Oscilloscope