NI PXIe-4154 Calibration Procedure
|
© National Instruments Corporation
|
3
Test Equipment
NI recommends that you use the equipment in Table 1 for calibrating the NI PXIe-4154.
Table 1.
Recommended Equipment
Equipment
Recommended
Model
Parameter
Measured
Minimum Requirements
Two Digital
Multimeters (DMM)
NI PXI-4071
All Parameters
Voltage: better than ±50 ppm
accuracy and better than 30 µV
resolution
100 m
Ω
Precision
Current Shunt
Guildline 9230A-15R
characterized at 5 W,
or Ohm Labs CS10
Current Output and
Measurement
±100 ppm tolerance.
±50 ppm stability.
±5 ppm/°C temperature
coefficient.
Minimum current 10 A.
4-wire kelvin sense.
100
Ω
Precision
Current Shunt
Vishay
Y1453100R000T9L
Current Output and
Measurement
±100 ppm tolerance.
±50 ppm stability.
±5 ppm/°C temperature
coefficient.
Minimum power rating 0.5 W.
External Current
Source
Fluke 5500A
or
Fluke 5520A
Current Measurement
Current output of at least 3 A.
Less than 8 µA RMS Normal
Mode Current Noise.
Programmable
Electronic Load with
Dynamic Current
Capability
Agilent N3302A
Load Regulation,
Transient Response,
Output Resistance
Constant Current mode with
the ability to sink at least 3 A.
Dynamic mode capable of
doing a 0.1 A to 1.5 A step with
a period of 5 mS and slew rate
of at least 200 mA/µS.
Digital Oscilloscope
NI PXI-5124
Transient Response
Sampling rate of at least
200 MS/s. Edge triggering
capability.
Two 22.6
Ω
Resistors
Vishay
PTF22R600FYBF
Remote Sense Output
1% tolerance, ¼ W,
±10 ppm/°C
200
Ω
Resistor
Vishay
PTF65200R00AZEB
Remote Sense Output
0.05% tolerance, ¼ W,
±10 ppm/°C
Two 3-Foot, 18 AWG
Cables
Belden 8760-BEL
Transient Response
18 AWG, twisted-pair
Front Panel
Connector
Phoenix Contact
Combicon, 5.08 mm,
5 Position Connector
and Backshell Kit
NI P/N 781627-01
All Parameters
—