Trouble shooting
94
ENG_Manual_NW30-R01-01
Sterilizer NEWSTER
NW30
5.6.
Steam escapes from the cell
The output conduit of the vapors from the cell must remain free so that the suction due to the
fan of the filters group maintains the cell in slight depression. In this way, the vapors can not
get out of the cell by any other way except through the filter group.
If during the cycle the depression fails, a leak of vapours from the seal of the cell may occur.
It is possible that excessive dust has clogged the circular filter on the lid; in this case it is
necessary to open the cell and thoroughly clean the filter and then check that the vapor
suction hole is not blocked.
Fig. 82
If even after this intervention the steam continues to exit from the cell, put the machine in
"Manual" mode and turn on the vacuum cleaner filters in the group, cut a strip of paper 4 cm
wide and 20 long and bring it close to the vapour suction hole holding it tightly ; if the
paper is not sucked into the tube it is sure thar the columns of lowering temperatures are too
clogged with limestone. In this case the intervention of specialized technicians is required ,
please contact support.
5.7.
Table of the main problems, their causes and solutions.
PROBLEM
PROBABLE CAUSES
SOLUTION
Loud squeal coming from
the transmission group
Belt not stretched
Tighten the belts and check
their condition, change the
whole group, if necessary
(MAINTENANCE SPECIALIST)
Abnormal vibrations of the
machine running in jolts
The rotating blades are not
sharp enough or are too
rounded on the outside
toward the wall
Sharpen or replace the blades,
especially if very rounded
Excessive load of waste
Complete the cycle. In the future,
stick to a lighter load
The time required to cool
the treated material is
excessively long
The nozzles that send water
to the cell are blocked
Clean the nozzles
Vapors suction
hole
Summary of Contents for NW30
Page 1: ...ITA_MANUALE_NW30 R01 01 Sterilizzatore di rifiuti sanitari a rischio infettivo Manuale NW30 ...
Page 19: ...Sezione sicurezza 19 ITA_MANUALE_NW30 R01 01 Sterilizzatore NEWSTER NW50 2 Sezione sicurezza ...
Page 45: ...Sezione funzionamento 45 ITA_MANUALE_NW30 R01 01 Sterilizzatore NEWSTER NW30 Fig 27 ...
Page 90: ...Sezione manutenzione 90 ITA_MANUALE_NW30 R01 01 Sterilizzatore NEWSTER NW30 Fig 54 ...
Page 92: ...Sezione manutenzione 92 ITA_MANUALE_NW30 R01 01 Sterilizzatore NEWSTER NW30 Fig 55 ...
Page 94: ...Sezione manutenzione 94 ITA_MANUALE_NW30 R01 01 Sterilizzatore NEWSTER NW30 Fig 56 ...
Page 125: ...ENG_Manual_NW30 R01 01 Healthcare risk waste sterilizer User Manual NW30 ...
Page 129: ...General information 5 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW50S 1 General information ...
Page 140: ...Safety notes 16 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW50S 2 Safety notes ...
Page 144: ...Safety notes 20 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW30 Fig 17 ...
Page 159: ...Instructions for the user 35 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW30 Fig 26 ...
Page 188: ...Maintenance 64 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW30 4 Maintenance ...
Page 191: ...Maintenance 67 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW30 Fig 55 ...
Page 193: ...Maintenance 69 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW30 Fig 56 ...
Page 195: ...Maintenance 71 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW30 Fig 57 ...
Page 215: ...Trouble shooting 91 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW50S 5 Trouble shooting ...