Check the level of the reference electrolyte. It should be about 1cm below the filling
orifice, which is closed with a rubber T stopper. To add reference electrolyte, take the
filling pipette (P0740-4820) and fill it with Viscolyte B (P0860-0130) Electrolyte.
NOTE:
The two chambers are filled with same Reference Electrolyte.
The total volume of Reference Electrolyte held be the electrode
is approximately 30mL. During normal operation the two rubber
stoppers are to be removed.
3.6
pH ELECTRODE PROBE CALIBRATION
pH probe is to be calibrated prior to autoclaving vessel.
Connect electrode to probe cable.
Turn the agitation switch OFF.
Turn ON power switch.
NOTE:
The pH measuring system is calibrated using two buffer
solutions of a known pH.
1.
Set the selector switch to pH.
Set the mode switch to "ZERO".
Immerse the pH probe into an external pH 7.00 buffer solution.
Set the display to read the pH value of the buffer with INC/DEC switch.
2.
Immerse the pH electrode in a second external buffer solution which is several
pH units above or below the pH selected in the previous step.
Set the mode switch to "SPAN".
Set the display to read the value of the second buffer solution with the INC/
DEC
switch.
NOTE:
After autoclaving it is recommended that the calibration is
checked by taking a sample from the vessel and measuring
the pH with an external pH meter.
M1226-0050 12
Summary of Contents for BIOFLO III
Page 2: ......
Page 7: ......
Page 9: ...M1226 0050 2...
Page 37: ...M1226 0050 30...
Page 40: ...M1226 0050 33...
Page 48: ...FIGURE 1 VESSEL SCHEMATIC M1226 0050 41...
Page 49: ...FIGURE 2 SAMPLING SYSTEM M1226 0050 42...
Page 50: ...FIGURE 3 FLOWMETER CONVERSION M1226 0050 43...
Page 51: ...FIGURE 4 FRONT PANEL M1226 0050 44...
Page 52: ...FIGURE 5 REAR OF CABINET M1226 0050 45...
Page 53: ...FIGURE 6 SWITCH PANEL M1226 0050 46...
Page 54: ...FIGURE 7 EXTERNAL NUTRIENT PUMP TERMINAL M1226 0050 47...
Page 55: ...FIGURE 8 SUGGESTED IMPELLER LOCATION M1226 0050 48...
Page 56: ...FIGURE 9 HEADPLATE M1226 0050 49...
Page 57: ...FIGURE 10 SPARGER TUBE M1226 0050 50...
Page 58: ...FIGURE 11 HARVEST TUBE M1226 0050 51...
Page 59: ...FIGURE 12 THERMOWELL M1226 0050 52...
Page 60: ...FIGURE 13 FOAM PROBE M1226 0050 53...
Page 61: ...FIGURE 14 BAFFLE INSTALLATION M1226 0050 54...
Page 62: ...FIGURE 15 SIDE PORTS M1226 0050 55...
Page 63: ...FIGURE 16 pH PROBE INSTALLATION M1226 0050 56...
Page 64: ...FIGURE 17 D O PROBE PHOENIX INSTALLATION M1226 0050 57...
Page 65: ...FIGURE 18 D O PROBE INGOLD INSTALLATION M1226 0050 58...
Page 66: ...FIGURE 19 CONVERSION OF D O PORT TO INOCULATION PORT M1226 0050 59...
Page 67: ...FIGURE 20 EXHAUST CONDENSER INSTALLATION M1226 0050 60...
Page 68: ...FIGURE 21 RESTRICTOR IN CONDENSER WATER IN LINE M1226 0050 61...
Page 69: ...FIGURE 22 DISASSEMBLY OF CONDENSER M1226 0050 62...
Page 70: ...FIGURE 23 FACTOR AFFECTING pH CONTROL M1226 0050 63...
Page 86: ...M1226 0050 79...
Page 87: ...FIGURE 25 BEARING HOUSING ASSEMBLY M1226 0050 80...
Page 88: ...FIGURE 26 BEARING HOUSING IN STERILIZATION M1226 0050 81...
Page 89: ...M1226 0050 82...
Page 90: ...M1226 0050 83...
Page 91: ...M1226 0050 84...
Page 92: ...M1226 0050 85...