![NEW BRUNSWICK SCIENTIFIC BIOFLO III Manual To Operations Download Page 16](http://html1.mh-extra.com/html/new-brunswick-scientific/bioflo-iii/bioflo-iii_manual-to-operations_1677096016.webp)
before inserting into the port in the headplate.****
12.
Insert exhaust condenser into the exhaust condenser port. Connect the
exhaust filter on the top of the condenser with flexible tubing (Fig. 20).
13.
Slide 2" long, 1/4" inside diameter silicone tubing on the top of the sparger
tube, then connect air filter to it (Fig. 10).
NOTES:
*
Add a few drops of glycerin into the thermowell tube before inserting
RTD.
**
Do not overtighten the clamping ring to the headplate. Same to the
clamping screws for the glass tube reactor and the vessel base
(vessel
heater).
***
When the inoculation port is used for installation of Phoenix DO
probe, the DO port is converted to the inoculation port (Fig. 19).
**** Membrane of the probe must be free of glycerin.
3.3
PREPARATION BEFORE STERILIZATION
Before sterilization, perform the following procedure:
1.
Fill vessel with water to its working volume.
2.
Place the vessel assembly on the three pins located in the base of the cabinet.
3.
Position the motor assembly on the top of the bearing housing. Connect the
motor cable to the receptacle on the face of the cabinet.
4.
Connect cables from all probes (pH, DO and antifoam) to their respective
sockets on the front face of the cabinet.
Ground lead from the antifoam socket is to be connected to the ground pin
on
the
headplate.
5.
Insert RTD in the thermowell on the headplate and connect it to the socket
on
the
cabinet.
6.
Connect the inlet filter on the top of the sparger tube to the hose barb in the
face of the cabinet with flexible tubing.
7.
Connect water line to rear of the cabinet. Adjust water pressure to 20 PSIG
max.
M1226-0050 9
Summary of Contents for BIOFLO III
Page 2: ......
Page 7: ......
Page 9: ...M1226 0050 2...
Page 37: ...M1226 0050 30...
Page 40: ...M1226 0050 33...
Page 48: ...FIGURE 1 VESSEL SCHEMATIC M1226 0050 41...
Page 49: ...FIGURE 2 SAMPLING SYSTEM M1226 0050 42...
Page 50: ...FIGURE 3 FLOWMETER CONVERSION M1226 0050 43...
Page 51: ...FIGURE 4 FRONT PANEL M1226 0050 44...
Page 52: ...FIGURE 5 REAR OF CABINET M1226 0050 45...
Page 53: ...FIGURE 6 SWITCH PANEL M1226 0050 46...
Page 54: ...FIGURE 7 EXTERNAL NUTRIENT PUMP TERMINAL M1226 0050 47...
Page 55: ...FIGURE 8 SUGGESTED IMPELLER LOCATION M1226 0050 48...
Page 56: ...FIGURE 9 HEADPLATE M1226 0050 49...
Page 57: ...FIGURE 10 SPARGER TUBE M1226 0050 50...
Page 58: ...FIGURE 11 HARVEST TUBE M1226 0050 51...
Page 59: ...FIGURE 12 THERMOWELL M1226 0050 52...
Page 60: ...FIGURE 13 FOAM PROBE M1226 0050 53...
Page 61: ...FIGURE 14 BAFFLE INSTALLATION M1226 0050 54...
Page 62: ...FIGURE 15 SIDE PORTS M1226 0050 55...
Page 63: ...FIGURE 16 pH PROBE INSTALLATION M1226 0050 56...
Page 64: ...FIGURE 17 D O PROBE PHOENIX INSTALLATION M1226 0050 57...
Page 65: ...FIGURE 18 D O PROBE INGOLD INSTALLATION M1226 0050 58...
Page 66: ...FIGURE 19 CONVERSION OF D O PORT TO INOCULATION PORT M1226 0050 59...
Page 67: ...FIGURE 20 EXHAUST CONDENSER INSTALLATION M1226 0050 60...
Page 68: ...FIGURE 21 RESTRICTOR IN CONDENSER WATER IN LINE M1226 0050 61...
Page 69: ...FIGURE 22 DISASSEMBLY OF CONDENSER M1226 0050 62...
Page 70: ...FIGURE 23 FACTOR AFFECTING pH CONTROL M1226 0050 63...
Page 86: ...M1226 0050 79...
Page 87: ...FIGURE 25 BEARING HOUSING ASSEMBLY M1226 0050 80...
Page 88: ...FIGURE 26 BEARING HOUSING IN STERILIZATION M1226 0050 81...
Page 89: ...M1226 0050 82...
Page 90: ...M1226 0050 83...
Page 91: ...M1226 0050 84...
Page 92: ...M1226 0050 85...