Appendix
LFA 447
TM
01/09 J:\LFA447\CHAPTER5.DOC
12
Reassembly to the 300°C-furnace
The reassembly to the 300°C furnace with
four samples must be carried out in reverse
order.
ATTENTION!
•
sensor:
sensor
guide pins
For the 300°C furnace the sensor must be
placed in the lower position so that the guide
pins go through the holes in the sensor plate
as shown in the figure.
•
Software settings:
software settings for the 300°C furnace:
- The standard sample holder with
Four
Samples
must be selected in the instrument
configuration:
Menu:
File
Menu item:
Instrument Configuration
Sample Tray
→
Four
Samples
- The furnace type
300
must be
selected in the instrument configuration:
Menu:
File
Menu item:
Instrument Configuration
Furnace Type
→
300