Appendix
LFA 447
TM
01/09 J:\LFA447\CHAPTER5.DOC
7
sensor (upper position)
1
2
16.
Insert the sensor and connect the
plug.
ATTENTION!
When using the Matrix sample
holder, the sensor must be placed in
the
upper position
contrary to the
normal sample holder.
Lift the sensor from the normal
position upwards and turn it
counterclockwise until it snaps into
the upper end position.
(In the normal sensor position, the
guide pins go through the holes in
the sensor support. If the sensor is
placed in upper position, the guide
pins must snap into the sensor
support.)
guide pins
plug connection
thermocouple
furnace
a
b
a
b
17.
Disconnect the plug connection:
(only for 300°C furnace)
a – thermocouple
b – furnace
cable run
18.
Remove the cable from the cable
run.