System Components
LFA 447
TM
01/09 J:\LFA447\CHAPTER3.DOC
11
•
Sample holder for liquids
By using a special sample holder it is possible to analyze liquids with the LFA 447 instrument.
When using the sample holder please pay attention to the following indications.
lid
crucible
liquid sample
ring
of stainless steel
Figure 12: sample holder for liquids
Preparing the sample holder
The bottom side of the crucible and the indentation on the top side of the lid must be
coated with graphite spray as shown in Figure 13.
lid
(coating)
crucible
(bottom side,
with coating)
ring of
stainless steel
Figure 13: coatings on the cover and crucible
If the upper edge of the lid is soiled when applying the graphite spray, it must be
cleaned with alcohol.