![NETZSCH LFA 447 Nanoflash Operating Instructions Manual Download Page 41](http://html1.mh-extra.com/html/netzsch/lfa-447-nanoflash/lfa-447-nanoflash_operating-instructions-manual_1675675041.webp)
System Components
LFA 447
TM
01/09 J:\LFA447\CHAPTER3.DOC
15
extractor tool
furnace lid
•
Put the furnace lid in the place.
A test measurement with water should yield the following result:
a = 0.146 mm
2
/s (at room temperature)
Example of a measurement:
Liquid sample holder
304 V / Long / 3500ms*
Signal Ratio (empty/water) < 1/7
3 shots / calc. Range 2000ms:
a_water_25°C = 0.148 mm²/s (Lit: 0.146 mm²/s; +1.4%)
* A calculation range of max. 2000 ms is recommended for all liquids.