System Components
LFA 447
TM
01/09 J:\LFA447\CHAPTER3.DOC
9
In addition:
The sample holder for laminated samples can also be used for in-plane measurements
(anisotropic samples). In order to achieve this, the sample must also be cut into single
sample strips (E) (strips length: 12.7 mm) which must be turned by 90° (on their side) (F)
so that the sample can be measured in a horizontal direction. The sample strips must be
placed into the sample holder plate (G), affixed by the two fixing screws (H) and covered
by the corresponding mask (I) (see Figure 10).
sample
operating principle:
sample strip
sample strips
mask
fixing screws
fixing screws
sample holder plate
E
G
F
H
I
90°
square sample
sample strip
sample strip
Figure 10: operating principle sample holder for laminated samples for in-plane measurements