Operating the Instrument
LFA 447
TM
01/09 J:\LFA447\CHAPTER4.doc
20
Procedure for cp-determination
The determination of the specific heat using LFA 447 occurs via the comparison method. Exactly
the same test conditions for sample and reference are absolutely necessary.
Please note the following indications:
Right choice of the reference sample
-
Reference samples are offered for each material group: metals (e.g. SRM1461); polymers,
ceramics (e.g. Pyroceram); C-composites (Poco graphite).
-
Sample and reference should show similarly the same surface structure.
Right choice of the sample geometry
-
Ideally the dimensions of the sample and reference should be equal
→
different
dimensions are offered for all reference samples.
-
Different diameter or edge lengths do mostly not cause an negative effect on the
measuring results as long as the whole bottom side of the sample is lighted and the same
aperture diameter is used (the aperture diameter must always be smaller than the
diameter of the sample plate, ratio 0.7 to 0.8).
-
Increase variations result if the thickness of the sample and reference is strong different,
But: important is the distance between the surface of the sample and IR-detector, viz:
through balancing of the difference of level good measuring results can be achieved, it is
recommended to balancing the difference of level from thickness differences of more
than 0.5 mm.
Correct coating
-
The coating with graphite spray is used on the one hand for an increase of the emission
ratio and the absorptance of the samples so that the signal-to-noise ratio is improved
and otherwise to a homogenization of the surface (similar and homogenize energy input
for sample and reference).
-
Coat 2 to 3 thin layers (Shake the can and hold it approximately 30 cm from the samples),
clean the sample with alcohol – This must be done before each measurement!
Other checks
-
Check if the windows are clean before each measurement (Borofloat-glass between
furnace and flash lamp as well as the 4 sapphire windows of the furnace lid).
-
Check that nothing stands between the flash lamp and the sample (connecting cable or
similar, especial for sample position 1 and 3), if necessary remove the front cover, take
out the Borofloat-glass, loose the cable connection and correct the position.
Summary of Contents for LFA 447 Nanoflash
Page 2: ...01 09 J LFA447 CHAPTER1 DOC Chapter I General Information LFA 447TM...
Page 18: ...01 09 J LFA447 CHAPTER2 DOC Chapter II Installation LFA 447TM...
Page 25: ...01 09 J LFA447 CHAPTER3 DOC Chapter III System Components LFA 447TM...
Page 50: ...01 09 J LFA447 CHAPTER4 doc Chapter IV Operating the Instrument LFA 447TM...
Page 70: ...Operating the Instrument LFA 447TM 01 09 J LFA447 CHAPTER4 doc 19...
Page 73: ...01 09 J LFA447 CHAPTER5 DOC Chapter V Appendix LFA 447TM...