System Components
LFA 447
TM
01/09 J:\LFA447\CHAPTER3.DOC
16
The figure (example for water) shows that the calculation range is reduced to 2000 ms.
This is strongly recommended for all measurements with the sample holder for liquids
due to the heat spreading in the container which influences the detector signal above
2000 ms. In dependence on the thermophysical properties of the sample it can be
necessary to reduce the calculation range to a value between the maximum and
2000 ms.
-> Shots -> Set range for calculation -> Input values -> 2000 ms -> activate "all shots”
-> OK