System Components
LFA 447
TM
01/09 J:\LFA447\CHAPTER3.DOC
17
•
Sample holder for powdery and pasty samples
By using a special sample holder it is possible to analyze powdery and pasty samples with the
LFA 447 instrument. When using the sample holder please pay attention to the following
indications.
lower support plate
lid
pan
upper support plate
Figure 14: sample holder for powdery and pasty samples
Sample preparation for non-transparent samples:
pan
(bottom side)
graphite spary
cover plate
schematic:
pan
(bottom side)
cover plate
•
Turn the pan and place a cover plate
(Ø approx. 10 mm) on the bottom
side. Coat the bottom side with
graphite spray so that only the edge
is coated.
NOTE!
Only the transparent edge of the
pan must be coated with graphite
spray in order to prevent radiation
heat transfer through the sapphire
parts.