19
SPECIFICATION
S METER:
3-1/2 Digit LED
display.
LEAKAGE CURRENT:
0-1999 Micro Amps.
All Current measurements are made through
the AAMI/ANSI ES1-1985 or IEC 601-1 test
load. The meter readings correspond to the
true RMS value of the current.
Accuracy:
±
1% of reading, + 1
LSD DC to 1 KHz,
±
2.5% of r 1 LSD
1 Khz to 100 KHz,
±
5% of r 1 LSD
100 KHz to 1 MHz,
RESISTANCE:
0 TO 1999 Milli Ohms.
Accuracy:
+ 1 % of r 1 LSD
DEVICE CURRENT:
0 to 19.99 Amps
Accuracy:
+ 2% Full Scale + 1 LSD
TEST RECEPTACLE:
Hospital Grade 110 VAC
-15 Amp or 220 VAC 10 Amp.
The rocker polarity switch selects NORMAL,
OFF and REVERSE Polarity to the test
receptacle, and a momentary Neutral switch will
open the neutral line to the test receptacle.
POWER REQUIREMENTS:
110 VAC 50-60Hz,
15 Amps or 220 VAC, 10 Amps maximum rated
for the test outlet and 0.1 Amp rated for the unit.