16
MAINTENANCE
The LKG 601 requires only minimum
maintenance.
Periodically check the inlet plug, outlet
receptacle, and wiring for any damage or wear
like cracks, cuts, or other defects. Also check
the case for any damage or cracks.
In order to ensure the accuracy, it is
recommended that the unit be checked
periodically. A calibrated output is provided at
the TEST JACK for Case Leakage and Cord
Resistance functions. This feature is useful to
check the integrity of the unit.
When making a resistance measurement with a
Test Cable other than the one supplied with the
LKG 601, the user can zero the leads by
adjusting the zero adjustment potentiometer.
This will not affect the internal calibration of the
unit.
Annual calibration is recommended. If the LKG
601 is returned to Netech for recalibration before
the first year warranty expires Netech will
provide a second year warranty. The warranty is
void if the LKG 601 is serviced by anyone other
than Netech or if the warranty seal is broken.