Troubleshooting
N-560 Service Manual
49
Data Port
Table 10 lists a symptom for data port problems and recommended actions. If
the action requires replacement of the User Interface PCB, refer to
Disassembly Guide
on page 53.
Error Codes
When the N
-
560 detects a defective sensor connected to the N-560, the N-560
displays an error code of “SEN Err.” The sensor should be replaced and the
N-560 power should be cycled.
An error code is displayed when the N-560 detects a non-correctable failure.
Cycling the power clears these errors. Table 11 lists the N-560 error codes
and their meanings. When one of these errors occur, the N-560:
• sounds a low priority alarm that cannot be silenced except by
power-down
• stops measurements
Table 10: Data Port Problems
Symptom
Recommended Action
No printout is being
received.
The N-560 is running on battery power. Connect the
N-560 to AC power.
The N-560's baud rate does not match the printer.
Change the baud rate of the N-560 following
instructions in the Operator’s manual.
Check the connection between the data port and the
printer. Refer to
Connecting To The Data Port
on
page 97.
Replace the Front Panel PCB. Refer to
Front Case
Disassembly
on page 57.
Summary of Contents for OXIMAX N-560
Page 1: ...C M Y CM MY CY CMY K Service ai 1 24 06 9 25 48 AM Service ai 1 24 06 9 25 48 AM...
Page 10: ...List of Tables N 560 Service Manual viii...
Page 38: ...Performance Verification 28...
Page 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Page 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Page 84: ...Spare Parts 74...
Page 88: ...Packing for Shipment 78...
Page 104: ...Specifications 94...
Page 114: ...Data Port Interface Protocol 104...
Page 140: ...Technical Discussion 142...
Page 144: ......
Page 145: ......