Data Port Interface Protocol
N-560 Service Manual
99
Nurse Call Polarity Settings
Nurse Call Relay Pin States
Nurse Call Relay Rating
Nurse Call Polarity
Alarm State
Voltage at pins
10 to 11
Normally High
No alarm or Alarms Silenced
5 to 12 VDC
Audible Alarm
-5 to -12 VDC
Normally Low
No alarm or Alarms Silenced
-5 to -12 VDC
Audible Alarm
5 to 12 VDC
Pin
No Alarm or
Alarm Silenced Audible Alarm
N-560 Turned
Off
7 N.O.
Open
Closed
Closed
8 N.C.
Closed
Open
Open
Maximum input voltage
30 VAC or 30 VDC
Load current
12 mA continuous (peak 300 mA at 100 msec)
Minimum resistance
26.5 ohms to 50 ohms (40.5 ohms typical during
alarms
Ground reference
isolated ground
Electrical isolation
1,500 V