Technical Discussion
N-560 Service Manual
107
Figure 46: Oxyhemoglobin Dissociation Curve
SatSeconds
Alarm Management
The N-560 utilizes Nellcor’s
SatSeconds
alarm management technique.
SatSeconds
is a function of the software within the N-560. With the
SatSeconds
technique, upper and lower alarm limits are set in the same way as
traditional alarm management. The clinician also sets a
SatSeconds
limit that
allows monitoring of %SpO
2
below the selected lower alarm limit for a period
of time before an audible alarm sounds. Refer to the N-560 Operator's manual
for managing
SatSeconds
.
Signal Processing
The N-560 takes advantage of increased microprocessing power with
advanced mathematical algorithms.
O
XI
M
AX
advanced signal processing
allows the N-560 to provide accurate saturation and pulse rate values, even
during difficult monitoring conditions of low perfusion and signal
interference.
Interference
is a term that was formally introduced in ISO9919.2005,
“Medical electrical equipment — Particular requirements for the basic safety
and essential performance of pulse oximeter equipment for medical use”.
Interference may arise from a number of sources, alone or in combination,
and the term refers to the effects on the incoming red and IR signals caused by
physiological conditions, medical procedures, or external agents. The effect
Summary of Contents for OXIMAX N-560
Page 1: ...C M Y CM MY CY CMY K Service ai 1 24 06 9 25 48 AM Service ai 1 24 06 9 25 48 AM...
Page 10: ...List of Tables N 560 Service Manual viii...
Page 38: ...Performance Verification 28...
Page 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Page 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Page 84: ...Spare Parts 74...
Page 88: ...Packing for Shipment 78...
Page 104: ...Specifications 94...
Page 114: ...Data Port Interface Protocol 104...
Page 140: ...Technical Discussion 142...
Page 144: ......
Page 145: ......