Data Port Interface Protocol
98
The pin layouts are illustrated in Figure 44. The conductive shell is used as
earth ground. An AMP connector is used to connect to the data port. Use
AMP connector (AMP P/N 747538-1), ferrule (AMP P/N 1-747579-2) and
compatible pins (AMP P/N 66570-2). The serial cable must have a braided
shield providing 100% coverage, such as Beldon cable (Beldon P/N 9616) or
equivalent. Connectors at both ends of the serial cable must have the shield
terminated to the full 360 degrees of the connector's metal shell.
CAUTION: Do not create sharp bends in the cable; this may tear or
break the shield.
Figure 44: Data Port Pin Layout
Pins 2, 3, and 5 provide data in RS
-
232 format.
Pins 1, 4, 9, and 12 provide data in RS-422 format.
10 Signal Ground
11 Nurse Call (RS-232 level output)
12 TXD- (RS-422 negative output) (-5 to -12 VDC with no audible
alarm, 5 to 12 VDC with audible alarm)
13 NC
14 NC
15 Nurse Call Common for Dry Contacts
Table 19: Data Port Pinouts (Continued)
Pin
Signal Name
Summary of Contents for OXIMAX N-560
Page 1: ...C M Y CM MY CY CMY K Service ai 1 24 06 9 25 48 AM Service ai 1 24 06 9 25 48 AM...
Page 10: ...List of Tables N 560 Service Manual viii...
Page 38: ...Performance Verification 28...
Page 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Page 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Page 84: ...Spare Parts 74...
Page 88: ...Packing for Shipment 78...
Page 104: ...Specifications 94...
Page 114: ...Data Port Interface Protocol 104...
Page 140: ...Technical Discussion 142...
Page 144: ......
Page 145: ......