Performance Verification
8
Performance Tests
Note
: This section is written using Nellcor factory-set defaults. If your
institution has preconfigured custom defaults, those values will be displayed.
Factory defaults can be restored. See
Menu Item 10 (Restore Factory Default
Settings as Power-On Default)
on page 38.
Battery Charge
Perform the following procedure to fully charge the battery.
1. Connect the N-560 to an AC power source.
2. Verify that the N-560 is off and that the
AC Power
indicator is lit.
3. Charge the battery for at least 6 hours in standby.
Power-Up Performance
The power-up performance tests verify the following N-560 functions:
•
Power-On Self-Test (POST)
on page 9
•
Power-On Defaults and Alarm Limit Ranges
on page 11