![Nellcor OXIMAX N-560 Service Manual Download Page 34](http://html2.mh-extra.com/html/nellcor/oximax-n-560/oximax-n-560_service-manual_3907265034.webp)
Performance Verification
24
• 75 %SpO2 (test pass criteria is 73 to 77 %SpO
2
)
• 60 bpm (test pass criteria is 57 to 63 bpm)
• alarm
•
Pulse Amplitude
indicator - high level modulation
4. Perform
Test #1: BPM
on page 21. The
Pulse Amplitude
indicator
should indicate high level modulation.
5. Perform
Test #2: SpO2
on page 22. The
Pulse Amplitude
indicator
should indicate high level modulation.
6. Press the SRC-MAX % Modulation selection button. The SRC-MAX %
Modulation
LED will light.
7. The N-560
Pulse Amplitude
indicator will stabilize and illuminate 4
bars with each pulse beat.
8. The N-560 will display:
• 75 %SpO2
• 60 bpm
• alarm
•
Pulse Amplitude
indicator - low level modulation
9. Perform
Test #1: BPM
on page 21. The
Pulse Amplitude
indicator
should indicate low level modulation.