Routine Maintenance
6
2. Perform the electrical safety tests detailed in
Safety Tests
on page 26. If
the N-560 fails these electrical safety tests, refer to
Troubleshooting
on
page 43.
3. Inspect the fuses for proper value and rating (F1 and F2 = 2 amp,
250 volts).
Battery
Nellcor recommends replacing the N-560's battery every two years. When the
N-560 is going to be stored for two months or more, remove the battery prior
to storage. To replace or remove the battery, refer to
Disassembly Guide
on
page 53.
If the N-560 has been stored for more than 30 days, charge the battery as
described in
Battery Charge
on page 8. A fully discharged battery requires
6 hours to receive a full charge. The battery is being charged whenever the
N-560 is plugged into an AC power source.
Summary of Contents for OXIMAX N-560
Page 1: ...C M Y CM MY CY CMY K Service ai 1 24 06 9 25 48 AM Service ai 1 24 06 9 25 48 AM...
Page 10: ...List of Tables N 560 Service Manual viii...
Page 38: ...Performance Verification 28...
Page 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Page 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Page 84: ...Spare Parts 74...
Page 88: ...Packing for Shipment 78...
Page 104: ...Specifications 94...
Page 114: ...Data Port Interface Protocol 104...
Page 140: ...Technical Discussion 142...
Page 144: ......