Performance Verification
N-560 Service Manual
21
2. Connect the SRC-MAX tester to the other end of the DOC-10 cable.
3. Turn on the N-560 by pressing the
Power On/Off
button.
4. After the N-560 completes POST, the N-560 will:
• be in SpO
2
alarm
• display an %SpO
2
of 75 (pass criteria is 73 to 77 %SpO
2
)
• display a pulse rate of 60 (pass criteria is 57 to 63 bpm)
•
Pulse Amplitude
indicator - display low level modulation (low
amplitude
Pulse Amplitude
indicator)
Test #1: BPM
1. Press the SRC-MAX % Pulse Rate selection button. The SRC-MAX
Pulse Rate 200 LED will light.
2. The N-560 bpm will increase to 200 and stabilize at 200 bpm. The test
pass criteria is 197 to 203 bpm.
3. The N-560 will display:
• 75 %SpO2
• 200 bpm (pass criteria is 197 to 203 bpm)
• alarm
•
Pulse Amplitude
indicator - low level modulation