Data Port Interface Protocol
96
4. Press the
SatSeconds
Alarm Limit button to select option 5.
5. Select protocol 1 or 2 by pressing
Adjust Up
or
Adjust Down
button.
6. Press the
SatSeconds
Alarm Limit button to set the selection.
Baud Rate
Menu item 4 is used to select baud rate. To access menu item 4:
1. Disconnect the sensor cable.
2. Press both the
SpO
2
Alarm Limit
and
Pulse Rate Alarm Limit
buttons
simultaneously for 3 seconds.
3. Press the
Adjust Up
button until menu item 4 is displayed.
4. Press the
SatSeconds
Alarm Limit button to select option 4.