Specifications
N-560 Service Manual
93
Patient Risk Applied Current
Patient Isolation Risk Current
AC Line
Polarity
Neutral Line
Cord
Power Line
Ground
Cord
IEC 60601
-
1,
UL 2601
-
1
AAMI/
ANSI
-
ES1
Normal
Closed
Closed
100 µA
10 µA
Normal
Open
Closed
500 µA
50 µA
Normal
Closed
Open
500 µA
50 µA
Reversed
Closed
Closed
100 µA
10 µA
Reversed
Open
Closed
500 µA
50 µA
Reversed
Closed
Open
500 µA
50 µA
AC Line
Polarity
Neutral Line
Cord
Power Line
Ground
Cord
IEC 60601
-
1,
UL 2601
-
1
AAMI/
ANSI
-
ES1
Normal
Closed
Closed
5 mA
50 µA
Reversed
Closed
Closed
5 mA
50 µA