Chapter 2 Routine Maintenance Procedures
F
IGURE
2-2
M
EASURING DISTANCE
Item
Description
Notes
6
Lens
8
Light intensity rocker switch
Sets intensity to high or low
9
Target light rocker switch
Controls red target light; momentary action
33
On/standby switch
Lights up when device is on
A
Radiometer
B
Distance
1
Determine the distance (B), which corresponds to how far the LED light panel will be from the patient
during therapy.
Important!
The light intensity was factory calibrated with the neoBLUE
®
Radiometer and was based off
the distance (B) of 12 inches (30.5 cm).
2
Position the stand to suspend the light panel over the work pad.
3
Place the radiometer on the work pad under the light panel. The distance (B) should be measured from
the work pad surface to the bottom of the lens (6) as shown in Figure 2-2.
4
Check the intensity on the HIGH setting:
•
Place the light intensity rocker switch (8) in the HIGH position.
•
Turn on the illuminated On/standby switch. Verify that the enclosure fan is running and that the
LED light panel lights up.
•
Power on the radiometer and place it in the approximate center of the work surface below the
UUT.
•
After a stable reading is attained the radiometer should be moved as necessary to establish a
peak light reading. This should be in the approximate center of the area beneath the UUT
diffuser.
A
6
8
33
B
9
18
neoBLUE LED Phototherapy System Service Manual
PN 001320 Rev K