Chapter 7 Electromagnetic Specifications
Chapter 7
Electromagnetic Specifications
The neoBLUE
®
LED Phototherapy device is intended for use in the electromagnetic environment
specified below. The customer or the user of the neoBLUE
®
LED Phototherapy device should assure that
it is used in such an environment.
Table 7.1
Electromagnetic emissions
Emissions Test
Compliance
Electromagnetic Environment — Guidance
Radio Frequency
(RF)
Emissions
CISPR 11
Group 1
The neoBLUE
®
LED Phototherapy device uses RF energy only for
its internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class A
The neoBLUE
®
LED Phototherapy device is suitable for use in all
establishments other than domestic and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic
Emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Complies
42
neoBLUE LED Phototherapy System Service Manual
PN 001320 Rev K
Summary of Contents for neoBLUE
Page 1: ...Service Manual...