4
If the AC power entry module is being replaced:
•
Obtain the replacement and reverse these steps to install.
•
Reverse all steps in Section 4.2 on page 28 until the UUR is completely assembled.
•
Return to Chapter 2 and perform the safety tests and performance verification and adjustment.
Important!
Ensure that the ground wire to plate is attached after either disassembling or re-assembling
the UUR.
4.3.3 Removing the switches
1
Disconnect the spade connectors from the switch to be removed. Take note of wire colors and
connections.
2
Remove the switch by compressing the spring fasteners on either side of the switch housing. The
switch can be pushed out the rear of the enclosure.
3
If a switch is being replaced:
•
Obtain the replacement and reverse these steps to install.
•
Reverse all steps in Section 4.2 on page 28 until the UUR is completely assembled.
•
Return to Chapter 2 and perform the safety tests and performance verification and adjustment.
4.3.4 Removing the timer
1
Press the side tabs of the plastic clamp, holding the timer in, and pull back to free it from the timer.
2
Disconnect the spade connectors from the timer.
3
Disconnect the cable from the J7 header on the current PCB.
4
If the timer is being replaced:
•
Obtain the replacement and reverse these steps to install.
•
Return to Chapter 2 and perform the safety tests and performance verification and adjustment.
•
Reverse all steps in Section 4.2 on page 28 until the UUR is completely assembled.
neoBLUE
®
LED Phototherapy System Service Manual
31
PN 001320 Rev K