Note:
The radiometer should read 35 ±3.5 µW/cm
2
/nm at 12 inches (30.5 cm) based on the factory
calibration. If the reading is not within the tolerance, note the value for use later in this
procedure.
5
Check the intensity on the LOW setting:
•
Place the light intensity rocker switch in the LOW position.
•
Obtain a peak reading as in Step 4.
Note:
The radiometer should read 15 ±2 µW/cm
2
/nm at 12 inches (30.5 cm) based on the factory
calibration. If the reading is not within the tolerance, note the value for use later in this
procedure.
6
If all criteria in the previous steps were met, the UUT performance has been verified and the UUT can
be returned to the user. Turn off the UUT and ignore all further steps in this chapter.
7
If light intensity on either setting was not within tolerance, leave the test setup as is and proceed to
Section 2.4 on page 22.
2.4 Adjusting light intensity
F
IGURE
2-3
A
DJUSTING THE POTENTIOMETERS
1
Place the light intensity rocker switch in the HIGH position.
•
Examine the radiometer reading.
•
Use the adjustment tool to adjust potentiometer A (in Figure 2-3) for a light intensity of
35
± 3.5
µW/cm
2
/nm.
2
Place the light intensity rocker switch in the LOW position.
•
Examine the photometer reading.
•
If necessary, use the adjustment tool to adjust potentiometer B for a light intensity of
15
±
2
µW/cm
2
/nm.
3
If the tolerances in steps 4 and 5 in Section 2.3.2 are met, the UUT can be returned to the user.
B
A
neoBLUE
®
LED Phototherapy System Service Manual
19
PN 001320 Rev K