Chapter 2 Routine Maintenance Procedures
2.1 Cleaning
Cleaning the UUT incorporates two different tasks:
•
Cleaning the light enclosure and accessories (Section 2.1.1)
•
Cleaning the light enclosure air filters
(Section 2.1.2)
WARNING
Disconnect UUT from AC power before cleaning.
WARNING
Biohazard. Until UUT is disinfected per institutional
requirements, gloves and other appropriate safeguards
must be employed.
CAUTION:
Do not immerse the UUT or its accessories in liquid, or clean with caustic or abrasive
cleaners. Do not spray or pour any liquid on the UUT enclosure.
2.1.1 Cleaning the light enclosure and accessories
Clean the UUT enclosure using a cloth dampened with Germiocide™ or an equivalent product. Wipe the
exterior plastic surfaces lightly. Do not allow any liquid to penetrate the power connector, switches, fans,
connectors, or openings on the enclosure.
2.1.2 Cleaning the light enclosure air filters
The air filters should be cleaned at least every 30 days to prevent possible instrument overheating.
1
Remove the covers from the enclosure fan filter assemblies, shown below in Figure 2-1.
2
Inspect filters for damage. If damaged, they must be replaced.
1
Clean the filter by holding it under warm running water to clean away dust and other foreign materials.
F
IGURE
2-1
L
OCATING THE TWO FILTERS ON BACKSIDE OF ENCLOSURE
14
neoBLUE LED Phototherapy System Service Manual
PN 001320 Rev K