![National Instruments NI PXIe-5450 User Manual Download Page 88](http://html2.mh-extra.com/html/national-instruments/ni-pxie-5450/ni-pxie-5450_user-manual_3486740088.webp)
Chapter 2
NI 5450 Overview
©
National Instruments Corporation
2-65
NI PXIe-5450 User Manual
process. The size of the waveform you wish to stream must be evenly
divisible by the amount of onboard memory allocated for streaming to
prevent the streaming waveform from being overwritten before it has
generated.
2.
Identify the streaming waveform
—Set the Streaming Waveform
Handle property or the
NIFGEN_ATTR_STREAMING_WAVEFORM_HANDLE
attribute to the
waveform handle returned in step 1. Setting this property or attribute
ensures that none of your streaming data is overwritten before it is
generated. NI-FGEN monitors your progress to ensure that you write
fresh data fast enough to keep up with the generation. If your
application fails to keep up or attempts to write fresh data over data that
has not been generated, NI-FGEN returns an error.
1.6 GB Waveform
160 MB
Remaining Onboard
Memory Used for
Additional Waveforms
and Script Instructions.
160 MB Allocated
for Streaming
1.6 GB Waveform
160 MB
Set the Streaming Waveform Handle Property
to Identify the Waveform for Streaming