Chapter 2
NI 5450 Overview
©
National Instruments Corporation
2-89
NI PXIe-5450 User Manual
Synchronization
Your signal generator is based on the National Instruments
Synchronization and Memory Core (SMC) technology and therefore
supports TClk synchronization. Refer to the NI-TClk Synchronization
Help for more information about NI-TClk Synchronization.
Specifications
For information about the signal generator specifications, refer to the
device specifications. You can access these specifications by navigating to
Start»All Programs»National Instruments»NI-FGEN»
Documentation»Hardware Specifications
, or you can visit
ni.com/manuals
.
Calibration
Before shipping your signal generator, NI calibrated your device to ensure
that all features are within specifications.
Calibration is a set of operations that compares the values indicated by a
measuring instrument or measuring system to the corresponding values
realized by external standards. You can use the results of calibration to
determine the measurement error and can then correct for it in the
adjustment process.
The calibration process consists of verifying, adjusting, and reverifying a
device. During verification, you compare the measured performance to an
external standard of known measurement uncertainty to confirm that the
product meets or exceeds specifications. During adjustment, you correct
the measurement error of the device by adjusting the calibration constants
and storing the new calibration constants in the EEPROM. The host
computer reads the calibration constants and the software uses them to
compensate for errors in the data and to present calibrated data to the user.
Accessories
National Instruments offers a variety of products to use with your signal
generator, including cables and other accessories. Visit
for more
information.