![National Instruments NI PXIe-5450 User Manual Download Page 73](http://html2.mh-extra.com/html/national-instruments/ni-pxie-5450/ni-pxie-5450_user-manual_3486740073.webp)
Chapter 2
NI 5450 Overview
NI PXIe-5450 User Manual
2-50
ni.com
Total Onboard Memory Used
= 1,281,408 bytes
2.
An application requires using six waveforms with the following sizes:
480; 260; 960; 492; 516; and 604 samples. The waveforms are
generated by using Arbitrary Sequence mode and Burst trigger mode
to configure 10,000 segments in a sequence list. The following table
shows all the numbers used to determine the total memory stored in the
onboard memory: 135,296 bytes.
Total Onboard Memory Used
= 135,296 bytes
Number of
Segments in
Sequence
Memory
Calculation
Bytes
Rounded Size
20,000
80 +
(64 × 20,000) =
1,280,080
1,280,128
Waveforms
Samples
Bytes
Rounded Size
A
480
960
1,024
B
260
520
640
C
960
1,920
1,920
D
492
984
1,024
E
516
1,032
1,152
F
604
1,208
1,280
Memory Size =
7,040
Number of
Segments in
Sequence
Memory
Calculation
Bytes
Rounded Size
10,000
160 + (128 ×
10,000) =
128,160
128,256