ACCESS and ACTIVE LEDs
The ACCESS and ACTIVE LEDs indicate the status of the PXIe module:
ACCESS LED
The ACCESS LED indicates basic hardware status as shown in the
following table.
ACTIVE LED
The ACTIVE LED indicates the device state as shown in the following
table.
Color
Indications
Off
Device is not yet functional, or the device has
detected a problem with a power rail.
Amber
The device is being accessed.
Green
The device is ready to be programmed by
NI-FGEN.
Color
Indications
Off
Device is not generating.
Amber
The device is armed and waiting for a trigger.
Green
The device has received a trigger. Also indicates
that the device is generating a waveform.
Red
The device has detected an error. NI-FGEN must
access the device to determine the cause of the
error. The LED remains red until the error condition
is removed. Errors might include unlocked PLLs or
an over-temperature condition.