25. Replace the side panels on the test rack.
26. Secure the side panels by turning the four quarter-turn locks on each panel.
Configuration Steps
27. Reset the RMX power supply to the default factory settings.
a)
Flip the power switch to the on position.
b)
Press the MENU button to illuminate the menu LED and wait for
SEt
to appear on the voltage display.
c)
Turn the voltage encoder until
MEMO
appears.
d)
Press the voltage encoder and wait for
SAVE
to appear.
e)
Turn the voltage encoder until
F.rst
appears.
f)
Press the voltage encoder and wait for
F.rst
to appear on the voltage display and
YES
to appear on the current display.
g)
Press the current encoder and wait for
HOLD
to appear.
h)
Press the MENU button twice to turn the menu LED off.
28. Configure the power supply IP address.
a)
Press the REM button on the front panel of the power supply.
b)
Press the current encoder and wait for
Adr
to appear on the current display.
c)
Turn the current encoder to address
6
.
d)
Press the current encoder to save the address.
29. Reset the power supply to the LAN default factory settings.
a)
Press the REM button on the remote.
b)
Turn the voltage encoder until
rST
appears on the voltage display.
c)
Press the current encoder and wait for
rST
to appear on the current display.
d)
Press the current encoder to reset the system.
e)
Flip the power switch to the off position and leave the RMX-410 power supply off for five seconds.
f)
Flip the power switch to the on position.
30. Configure the static IP address.
a)
Press the REM button on the front panel of the power supply and wait for
IntF
to appear on the voltage encoder.
b)
Press the voltage encoder.
c)
Turn the current encoder until
LAn
appears.
d)
Press the current encoder.
e)
Press the REM button.
f)
Turn the voltage encoder until
IP
appears on the voltage display.
g)
Press the voltage encoder.
h)
Turn the voltage encoder to view the IP address.
Note
The voltage display will show IP1-IP4, where IP1 refers to the first section of the IP address. The current display
shows the required IP address.
i)
Turn the current encoder to change the IP address.
j)
Press the current encoder to select the IP address.
k)
through j for each IP address.
Note
Order the power supply IP address from left to right and bottom to top. Refer to
page 6 for more information on the assigned IP addresses for RMX power supplies.
31. Add the power supply to NI MAX.
a)
Sign in the ECUTS-16001.
b)
Launch NI MAX and navigate to
My System
»
Devices and Interfaces
»
Network Devices
.
c)
Right-click
Network Devices
.
d)
Select
Auto-detect of LAN Instrument
in the dialog box.
e)
Select the RMX power supply automatically detected by NI MAX.
f)
Click
Finish
.
g)
Click on the RMX power supply under
Network Devices
to show its configuration pane.
h)
Change the last
X
in
PPS_RMX_X
to match the final number in the IP address of the RMX power supplies in the ECUTS-16001. For
example, IP address 169.254.130.21 will be PPS_RMX_1.
i)
Repeat steps c through h for each power supply.
Replacing the PXI Chassis Rear Fans
Note
Use caution when removing the fan assembly to avoid damaging the internal wire harness.
Equipment
•
#2 Phillips screwdriver
ECUTS-16000/16001 User Manual
|
© National Instruments Corporation
|
57