Calibration
NI provides calibration services for instruments in the ECUTS that require calibration. NI recommends that you adhere to the calibration
interval for all supported instruments in your system. For the latest information about available calibration service levels for your instrument,
refer to
Table 9. Instrument Calibration Intervals
Instrument
Function
Recommended Calibration Interval
PXIe-4081
DMM
2 years
PXIe-4082
DMM
2 years
PXIe-4112
Programmable Power Supply
2 years
PXIe-4139
SMU
1 year
PXIe-4309
Analog Input
2 years
PXIe-4310
Analog Input
2 years
PXIe-4322
Analog Output
1 year
PXIe-6738
Analog Output
2 years
PXIe-5105
Oscilloscope
2 years
PXIe-5110
Oscilloscope
2 years
PXIe-5163
Oscilloscope
2 years
PXIe-5413
Waveform Generator
2 years
PXIe-5433
Waveform Generator
2 years
PXIe-6366
Multifunction I/O Module
2 years
PXIe-6386
Multifunction I/O Module
2 years
PXIe-6375
Current measurement for SLSC
2 years
USB-6363
Current measurement for SLSC
2 years
SLSC-12251
Current Measurement for SLSC
2 years
SLSC-12252
Current Measurement for SLSC
2 years
RMX-4003
Electronic Load
1 year
RMX-4005
Electronic Load
1 year
Factors such as shelf life and storage temperature can affect an instrument's recommended calibration interval. Refer to the specifications
document for your instrument to calcuate the appropriate calibration interval.
Troubleshooting
ECUTS Does Not Power On
•
Check the power source connection and voltage supply.
•
Verify that the main power switch is turned to the On position.
•
Confirm that the EPO button is not engaged.
•
Verify that the main breaker is in the On position on the PEP.
•
Verify that the temperature controller on the EPO panel has not detected a temperature higher than the programmed shutoff value.
•
Checking the incoming power lights on the rear side of the PDU.
•
Verify that the PDU main breaker is on and that the control breaker is depressed.
•
Check that the cables in and out of the PDU are connected.
•
Verify the shutoff thermostats have been reset if the ECUTS shut down due to thermal overload.
PXI Chassis Does Not Power On
•
Confirm that the EPO button is not engaged.
•
Check if the PXI controller is properly seated inside the PXI chassis.
•
Check the PXI chassis LED indicators.
•
Check if the chassis power supply cable is connected.