5.7 Activating the alarm function
Narda
NBM-550
57
5.7
Activating the alarm function
The alarm function allows you to enter a limit value. The
instrument outputs an audible signal and the Status LED
flashes red when this limit value is exceeded.
This function is useful for checking limit values or for
signaling an early warning of dangerous field strengths, for
example.
The possible alarm limits are:
•
Normal probes
Table 3
Setting ranges of the limit values (in 1 dB steps)
•
Shaped probes: 0.1% – 10,000% (in 1 dB steps)
Defining the alarm limits
1. Open the
Alarm Limit
function
(MAIN/MEASUREMENT SETTINGS/...).
2. Use the arrow keys
▲
/
▼
to change the value and press
the
OK
key to confirm the setting.
Activating the alarm function
1. Open the
Alarm Function
function
(MAIN/MEASUREMENT SETTINGS/...).
2. Use the arrow keys
▲
/
▼
to select
On
and press the
OK
key to confirm the setting.
The symbol
is shown in the display when the
alarm function is activated (see
Measuring range
Minimum
Maximum
V/m
100 mV/m
100 kV/m
A/m
250
μ
A/m
250 A/m
W/m
2
25
μ
W/m
2
25 MW//m
2
mW/cm
2
2.5 nW/cm
2
2.5 kW/cm
2
Summary of Contents for NBM-550
Page 10: ...Contents VIII NBM 550 Narda ...
Page 38: ...3 Preparing the NBM 550 for use 28 NBM 550 Narda ...
Page 52: ...4 Getting started 42 NBM 550 Narda ...
Page 88: ...6 Measurements with the NBM 550 and EHP 50 78 NBM 550 Narda ...
Page 110: ...9 PC software 100 NBM 550 Narda ...
Page 139: ...12 10 CE Declaration of Conformity Narda NBM 550 129 12 10 CE Declaration of Conformity ...
Page 140: ...12 Specifications 130 NBM 550 Narda ...
Page 142: ...12 Specifications 132 NBM 550 Narda ...
Page 146: ...13 Ordering information 136 NBM 550 Narda ...
Page 154: ...Index 144 NBM 550 Narda ...
Page 155: ......