3.4 Power supply EHP-50
Narda
NBM-550
23
3.4
Power supply EHP-50
The EHP-50 Analyzer is fitted with a fixed, non-
exchangeable rechargeable battery which provides at least
9 hours of operation when fully charged. The battery is
charged using the charger unit supplied. It takes
approximately 6 hours to recharge the battery if it is almost
completely discharged.
Note:
You cannot operate the EHP-50 with the charger unit
connected as this considerably affects the measurement
characteristics of the EHP-50.
Charging the battery
Starting the charging cycle:
✓
The AC line voltage must match the operating voltage of
the charger unit.
Connect the charger unit to the AC line first and then
connect it to the charger socket on the EHP-50
(Figure 2 13d).
The charging cycle starts and the
DATA
LED flashes
rapidly:
– Orange: battery charging
– Green: battery fully charged
Note:
You can charge the battery with the EHP-50 switched on or
off. If you connect the charger unit when a measurement is
in progress, the measurement will stop and the charging
cycle starts. The EHP-50 remains switched off when you
unplug the charger unit.
WARNING
Charging the battery with an unsuitable
charger unit
The battery can overheat, explode or ignite
spontaneously.
You must only use the charger unit supplied with the
device for charging the battery.
Summary of Contents for NBM-550
Page 10: ...Contents VIII NBM 550 Narda ...
Page 38: ...3 Preparing the NBM 550 for use 28 NBM 550 Narda ...
Page 52: ...4 Getting started 42 NBM 550 Narda ...
Page 88: ...6 Measurements with the NBM 550 and EHP 50 78 NBM 550 Narda ...
Page 110: ...9 PC software 100 NBM 550 Narda ...
Page 139: ...12 10 CE Declaration of Conformity Narda NBM 550 129 12 10 CE Declaration of Conformity ...
Page 140: ...12 Specifications 130 NBM 550 Narda ...
Page 142: ...12 Specifications 132 NBM 550 Narda ...
Page 146: ...13 Ordering information 136 NBM 550 Narda ...
Page 154: ...Index 144 NBM 550 Narda ...
Page 155: ......