5.3 Measuring the spatial average
Narda
NBM-550
49
Setting the averaging time
You can set the time used to form the average for
Average
and MaxAvg
display types in the range 4 s to 30 min in
2 second steps.
Setting the time and resolution
1. Open the
Averaging Time
function
(MAIN/MEASUREMENT SETTINGS/...).
2. Select the digit using the
<
/
>
function keys.
3. Change the value using the arrow keys
▲
/
▼
.
4. Press the
OK
key to confirm the settings.
5.3
Measuring the spatial average
You can determine the spatial average of the field strength
using the Spatial Average function. This function is used to
determine the exposure to electromagnetic radiation of the
human body, for example.
Two measurement methods are available:
•
Discrete
Individual results are recorded and averaged. You can
measure at specific locations using this method.
•
Continuous
Results are recorded and averaged continuously during
the time that the probe is moved around the area of
interest. You can measure the field strength affecting a
particular space using this method.
Selecting the measurement method
1. Open the
Spatial AVG Mode
function
(MAIN/MEASUREMENT SETTINGS/Next/...).
2. Use the arrow keys
▲
/
▼
to select the method and press
the
OK
key.
Summary of Contents for NBM-550
Page 10: ...Contents VIII NBM 550 Narda ...
Page 38: ...3 Preparing the NBM 550 for use 28 NBM 550 Narda ...
Page 52: ...4 Getting started 42 NBM 550 Narda ...
Page 88: ...6 Measurements with the NBM 550 and EHP 50 78 NBM 550 Narda ...
Page 110: ...9 PC software 100 NBM 550 Narda ...
Page 139: ...12 10 CE Declaration of Conformity Narda NBM 550 129 12 10 CE Declaration of Conformity ...
Page 140: ...12 Specifications 130 NBM 550 Narda ...
Page 142: ...12 Specifications 132 NBM 550 Narda ...
Page 146: ...13 Ordering information 136 NBM 550 Narda ...
Page 154: ...Index 144 NBM 550 Narda ...
Page 155: ......