10 Time Domain (Scope) Mode
188
IDA
-
3106
Narda
10.4.2 Example 2: DECT telephone
DECT is characterized by the following properties:
• Frequency range: 1880 MHz – 1900 MHz
• Base station is permanently active
• Number of channels: 10
• Channel bandwidth: 1728 kHz
• Cycle time: 24 timeslots of 417 µs each = 10 ms
This results in the following settings:
Parameter
Function / Setting
Fcent
e. g. Channel 0:
1897.344 MHz
CBW
1728 kHz –> selected:
2 MHz
Measurement
High Resolution Scope
Time Span
24 timeslots x 417
s =
10 ms
Time Resolution
Automatically set from 1/CBW =
500 ns