179
The maximum values are also shown in addition to the actual or mean
values:
Table 32: Values displayed in
Level Meter
mode
Type
Function / Explanation
Peak
Displays the instantaneous measured value.
Note:
The peak value display (Peak) is held for 480 ms. The value
displayed therefore corresponds to the highest level measured
within the last 480 ms. This means that the numerical display is
more stable and the graphical display is smoothed.
Max
Displays the maximum value that has occurred since the
measurement started.
RMS
Displays the RMS value averaged over a defined period of time.
The bar next to
AVG
in the lower status bar indicates the progress
of the measurement until the averaging time is reached. When this
time is reached (complete bar), the instrument refreshes the
resulting value using the new measurement results and discards the
oldest results (
progressive
averaging over time).
RMS Max
Displays the maximum value of the averaged values that has
occurred since the measurement started.
Summary of Contents for IDA
Page 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Page 8: ...Contents VI IDA 3106 Narda...
Page 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Page 134: ...6 Common Functions 126 IDA 3106 Narda...
Page 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Page 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Page 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Page 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Page 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Page 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Page 246: ...15 PC Software 238 IDA 3106 Narda...
Page 254: ...16 Remote Control 246 IDA 3106 Narda...
Page 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Page 272: ...17 Specifications 264 IDA 3106 Narda...
Page 284: ...19 Glossary 276 IDA 3106 Narda...
Page 290: ...Index 282 IDA 3106 Narda...
Page 291: ......