7 Direction Finding Mode
130
IDA
-
3106
Narda
Table 31: Values displayed in
Direction Finding
mode
Type
Function / Explanation
Peak
Displays the instantaneous measured value.
The peak value display (Peak) is held for 480 ms. The value displayed therefore
corresponds to the highest level measured within the last 480 ms. This means that
the numerical display is more stable and the graphical display is smoothed.
RMS
Displays the RMS value averaged over a defined period of time. This setting should
be used for constant or fluctuating signals.
The bar next to
AVG
in the lower status bar indicates the progress of the
measurement until the averaging time is reached. When this time is reached
(complete bar), the instrument refreshes the resulting value using the new
measurement results and discards the oldest results (
progressive
averaging over
time). Please note that correct direction finding is only possible using a longer
averaging time if the direction of the antenna is also changed correspondingly
slowly.
Max
The maximum value (RMS or Peak) that occurred since the start of the
measurement is also shown. The maximum value is shown as a red line in the bar
graph. This display is only possible for
(see section 7.3 on
Summary of Contents for IDA
Page 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Page 8: ...Contents VI IDA 3106 Narda...
Page 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Page 134: ...6 Common Functions 126 IDA 3106 Narda...
Page 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Page 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Page 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Page 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Page 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Page 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Page 246: ...15 PC Software 238 IDA 3106 Narda...
Page 254: ...16 Remote Control 246 IDA 3106 Narda...
Page 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Page 272: ...17 Specifications 264 IDA 3106 Narda...
Page 284: ...19 Glossary 276 IDA 3106 Narda...
Page 290: ...Index 282 IDA 3106 Narda...
Page 291: ......