![NARDA 8718B User Manual Download Page 42](http://html1.mh-extra.com/html/narda/8718b/8718b_user-manual_3302313042.webp)
41
Chapter 7
Spatially-Averaged Measurements
6. Press
F1 (START)
while slowly moving the probe
vertically at a constant rate of movement.
7. Strive for an interval of about ten seconds between
starting and stopping the timer. The meter will beep
every second as a timing aid.
8. When the probe reaches the height equal to the top of
an adult’s head ( about 6 ft./1.9 m), press
F1
again
(
STOP
).
9. The meter will show the average value on the second
line of the display and the maximum value on the left
side of the third line of the display. The bar graph will
be indicating the instantaneous field magnitude in real
time. The averaging interval will be shown on the right
side of the third line, which can be useful in perfecting
your spacial averaging technique.
2 4 .
%
S
D
T
3 9 . 5 M A X
C F
1 .
0
T
S
A R T
M A X
M E N
Q U I T
2
F1
F2
F3
F4
0
R
U
1 7 .
%
S
D
T
E R A G E
3 9 . 5 M A X
1 0 . 4
S E C
T
S
A R T C L R
M E N U
Q U I T
8
F1
F2
F3
F4
A V
1 2 .
%
S
D
T
3 9 . 5 M A X
C F 1 .
0
S
T O P
1
Fl
F2
F3
F4
0
It is important to
move the probe
vertically at a con-
stant rate to obtain
an accurate
spa-
tially averaged
measurement.
N
ote
The averaging time
resolution is approxi-
mately 1/4 second.
N
ote
Summary of Contents for 8718B
Page 9: ...8 Operating the Meter...
Page 37: ...36 Operating the Meter...
Page 71: ...70 Operating the Meter...
Page 75: ...74 Operating the Meter...
Page 79: ...78 Operating the Meter...
Page 87: ...86 Operating the Meter...
Page 104: ...103 Chapter 18 User s Software Main Menu Screen...
Page 108: ...107 Chapter 18 User s Software Probe Installation Screen...
Page 112: ...111 Chapter 18 User s Software Logged Survey Data Screen...
Page 124: ...123 Appendix B Theory of Operation Model 8718B Functional Block Diagram...
Page 129: ...128 Appendices...
Page 133: ...132 Appendices...