112
Operating the Meter
d) The fourth line shows the
Run Ref. Number
that was
entered by the user at the beginning of the survey, the
Record
number and total number of records in the file
and the
Date
and
Start Time
when the first data point
was taken.
2. The lower section of the window shows the measured data in
both tabular and graphical form.
a) The table on the left side of the screen displays up to
eighteen (18) data points at one time. The
Data Point
#
number will be a continuous sequence shown
sequentially in the time the data points were taken.
The
Field Strength
is shown in the unit-of-measure
used during the survey. Mode 2 data (a series of
spatially averaged data points) shows both the
Average
and
Max.
(or peak) value for every data point.
Mode 3 data (continuous logging mode) shows the
Time
that each data point was taken
b) The bar graph on the right side of the screen shows
the same data listed in the table. The title of the bar
graph is
Logged Data
for Mode 1 data,
Spatial
Average & Max.
for Mode 2 data, and
Continuous
Log
for Mode 3 data.
c) The range of the vertical axis of the graph is
automatically adjusted to accommodate the data. The
number of data points will also vary depending on the
data with a maximum of thirty (30) displayed at one
time.
d) The
Scroll Back and Scroll Forward
buttons are used
when there are more than sixteen data points to view.
Each time you click on one of these buttons you
advance or go back by sixteen data points.
e) The
New Record and Old Record
buttons allow you
to go back and forth between different records within
the same file. Each survey is a different record
identified by the
Run Ref. Number
that was entered by
the user at the beginning of the survey.
If you used the
PAUSE feature dur-
ing continuous log-
ging, the time shown
for each data point
will be incorrect for
all data points after
the pause feature
was used.
N
ote
Summary of Contents for 8718B
Page 9: ...8 Operating the Meter...
Page 37: ...36 Operating the Meter...
Page 71: ...70 Operating the Meter...
Page 75: ...74 Operating the Meter...
Page 79: ...78 Operating the Meter...
Page 87: ...86 Operating the Meter...
Page 104: ...103 Chapter 18 User s Software Main Menu Screen...
Page 108: ...107 Chapter 18 User s Software Probe Installation Screen...
Page 112: ...111 Chapter 18 User s Software Logged Survey Data Screen...
Page 124: ...123 Appendix B Theory of Operation Model 8718B Functional Block Diagram...
Page 129: ...128 Appendices...
Page 133: ...132 Appendices...