17
Chapter 4
Making a Measurement
Zeroing the Probe
The probe and meter must be “zeroed” together. It is important
that the probe is not exposed to radiated fields during the
zeroing process. The storage case is shielded and provides a
convenient “zero density” environment.
To zero the probe...
1. Place the probe in the storage case. Be careful not
to move the probe while zeroing takes place (on
probes with low frequency test points, you can create
static which will give inaccurate results).
2. When you are ready, press the
ENTER
key. The
screen will say
ZEROING
.
3. The meter will beep and the main measurement screen
will appear in about 15 seconds after zeroing is completed.
If you get a message that the probe could not be zeroed, the
most likely causes are:
•
The probe is not connected properly
. Check the
connector and retry.
•
The meter is set for a fiber optic input and you are
using the standard cable input (or vice versa)
. Turn the
meter off. Turn the meter back on and press
(F4) (MENU)
to access the menu system. Press
8
(
NEXT
) to get to the
Second Menu Screen
. Press
3 (CABLE/FO
) to the probe input status. Press
F1 (OK)
if it is correct or
F4
to select the alternate. Then press
Esc
to return to the
Opening Menu Screen
.
7. Proceed with the probe zeroing process.
A 8 7 2 2 D
S N X X X X X
3 0 0
H
- 5 0 G H z
z
k
E - F L D
3 0 0
%
S T D
S H A P E D
Y E S
N O
M E N U
F1
F2
F3
F4
Position the probe in
the case with one
test point pointed
downward and the
other two test points
above the foam.
This will insure that
the test points are
not touching the
foam.
N
ote
Summary of Contents for 8718B
Page 9: ...8 Operating the Meter...
Page 37: ...36 Operating the Meter...
Page 71: ...70 Operating the Meter...
Page 75: ...74 Operating the Meter...
Page 79: ...78 Operating the Meter...
Page 87: ...86 Operating the Meter...
Page 104: ...103 Chapter 18 User s Software Main Menu Screen...
Page 108: ...107 Chapter 18 User s Software Probe Installation Screen...
Page 112: ...111 Chapter 18 User s Software Logged Survey Data Screen...
Page 124: ...123 Appendix B Theory of Operation Model 8718B Functional Block Diagram...
Page 129: ...128 Appendices...
Page 133: ...132 Appendices...