GB
9010/20 LCD
Sensor data sheet Serie 47 K-HT
Part No.: 10048199
Terminal Block
9010/9020
Series 47 K-HT
Caution:
Before connecting the measuring head bridge current, check hardware
and software settings.
On the 9010/9020 LCD, the cable shield is applied on one side.
Sensor type:
Catalytic/passive/constant current supply [3 wire passive]
Connection data:
Bridge current:
280 mA
Maximum rated current:
350 mA
Maximum rated voltage:
3.0 V
Power requirement:
1.0 W [excluding cable lengths]
Connection cable:
3-lead, 80% shielded
Maximum loop resistance:
32 Ohm
Maximum cable length:
1,000 m [with 1.5 mm
2
lead cross-section]
Cable diameter:
5,4 mm
Permitted lead cross-section:
1.0 – 2.5 mm
2
Terminal box: HT11 [100 °C]
Part No.: D0715154
Conditions for
use:
Assembly:
Wall-mounted
Protection:
Explosion protection:
Ex II 2G EEx d IIC T3 [-40 °C to +160 °C]
Approvals:
INERIS 03 ATEX 0208
Temperature:
-40 °C to +160 °C
Humidity:
5 – 95 % rel. humidity; non-condensing
Pressure:
800 – 1,200 hPa
Weight/sensor:
230 g
Dimensions, dia. x H:
36 mm x 56 mm
Housing material:
Stainless steel
Further details in the Operating and Maintenance Instructions
[Part No.: 10052472]
Summary of Contents for 9010 LCD
Page 1: ...Operating Manual CONTROL UNIT MODELS 9010 and 9020 LCD Order No 10093590 02 ...
Page 2: ...MSA AUER GmbH Thiemannstrasse 1 D 12059 Berlin Germany MSA AUER GmbH All rights reserved ...
Page 84: ...Control Unit Models 9010 LCD and 9020 LCD 84 Appendix MSA GB ...
Page 87: ......
Page 88: ......
Page 89: ......
Page 94: ......
Page 98: ......
Page 99: ......