Control Unit Models 9010 LCD and 9020 LCD
61
GB
9010/20 LCD
Sensor data sheet D-7711 K-PR
Part No.: 10046889
Terminal Block
9010/9020
Sensor Type D-7711 K-PR
Caution:
Before connecting the measuring head bridge current, check hardware
and software settings.
On the 9010/9020 LCD, the cable shield is applied on one side.
Sensor type:
Catalytic/passive/constant current supply [3 wire passive]
Connection data:
Bridge current:
310 mA, for all measuring components
Maximum rated current:
300 mA
Power requirement:
approx. 0,6 W [excluding cable lengths]
Connection cable:
3-lead, 80 % shielded
Maximum loop resistance:
36 Ohm
Maximum cable length:
1,000 m [with 1.5 mm
2
lead cross-section]
Cable diameter:
8.5 mm
Lead cross-section:
1.5 mm
2
Terminal box:
EEx e [Part No.: D0715205]
Conditions for
use:
Assembly:
Wall-mounted
Protection:
IP 42 [housing] to DIN 400 50
Explosion protection:
Ex II 2G EEx d IIC T4
-40 °C
Ta
+100 °C
Approvals:
PTB 01 ATEX 1152 X N1
Temperature:
-20 °C to +55 °C [methane to +100 °C]
Humidity:
5–95% relative humidity; non-condensing
Atmospheric Pressure:
900 – 1,100 hPa
Weight:
approx. 0.72 kg
Dimensions W x D x H:
50 mm x 40 mm x 50 mm
Housing material:
Stainless Steel [DIN 1.4305, V2A]
Heat-up time:
60 s
Response time t90:
Methane
10 s Hydrogen
10 s
Propane
14 s Acetone
18 s
Ethanol
20 s Hydrogen
25 s
Ethane
20 s
Summary of Contents for 9010 LCD
Page 1: ...Operating Manual CONTROL UNIT MODELS 9010 and 9020 LCD Order No 10093590 02 ...
Page 2: ...MSA AUER GmbH Thiemannstrasse 1 D 12059 Berlin Germany MSA AUER GmbH All rights reserved ...
Page 84: ...Control Unit Models 9010 LCD and 9020 LCD 84 Appendix MSA GB ...
Page 87: ......
Page 88: ......
Page 89: ......
Page 94: ......
Page 98: ......
Page 99: ......